Medical Device Software Development: Process, Standards, and Integration

Medical Device Software Development: Process, Standards, and Integration

Medical device software can control medical equipment, support clinical decisions, or connect devices with healthcare systems. Depending on its purpose, it may run on a physical device or operate as a standalone application.

Unlike conventional software, medical device software may directly affect patient safety. Its development therefore requires a structured approach to risk management and testing, with regulatory requirements determined by the software’s intended purpose and classification.

This guide explains how medical device software is classified and developed, which standards apply, and what it takes to integrate medical devices with electronic health records (EHRs) and cloud platforms.

Key Takeaways

  • Medical device classification depends on the software’s intended medical purpose.
  • IEC 62304 defines software lifecycle processes, while regulatory requirements vary by market.
  • Testing must demonstrate that the software meets requirements and supports its intended use.
  • Reliable EHR integration requires more than transferring data between systems.

What Is Medical Device Software?

Medical device software is software designed to perform a medical function or support the operation of a medical device. It may control equipment, analyze patient data, or enable communication between devices and healthcare systems. Its regulatory status depends on its intended purpose and how the software is used.

Medical device software for patient monitoring

Software in a Medical Device (SiMD)

SiMD runs as part of a medical device and supports its operation. For example, software inside an infusion pump controls medication delivery and monitors device performance. Development focuses on reliable operation and safe interaction with hardware.

Software as a Medical Device (SaMD)

SaMD performs a medical function independently of dedicated medical hardware. An application that analyzes CT scans to detect abnormalities is one example. Its development requires particular attention to clinical performance and the reliability of diagnostic results.

Connectivity and Companion Software

Connectivity software links medical devices with mobile apps, cloud services, or electronic health records (EHRs). It may transfer readings or make device data available to clinicians.

The FDA uses the term Medical Device Data Systems (MDDS) for systems that transfer, store, convert, or display device data without controlling the connected equipment. Under FDA rules, software functions solely intended to transfer, store, convert formats, or display medical device data generally fall outside the definition of a medical device, provided they do not analyze or interpret the data or control the connected device.

Comparison of SiMD, SaMD, and Connectivity Software
Software type Definition Example Regulatory pathway Development focus
SiMD Software operating as part of a medical device Infusion pump controller Typically assessed within the device’s regulatory pathway Hardware integration and operational safety
SaMD Standalone software performing a medical function CT image analysis application FDA device pathway or EU MDR/IVDR, depending on intended use and classification Clinical performance and software validation
Connectivity software Software exchanging data between devices and healthcare systems EHR gateway or companion app May fall outside device regulation or be regulated according to its functions Interoperability, data integrity, and cybersecurity

Is Your Software a Medical Device?

What makes software a medical device is not how it works, but what it is intended to do. The same technology can be used in both medical and non-medical applications. For example, an AI algorithm that recommends workout routines is generally not a medical device.

Medical device software classification and assessment

An algorithm that analyzes patient data to diagnose a disease, however, may qualify as one. Even a basic calculator can qualify as a medical device if its intended purpose is to guide treatment decisions.

  • According to the FDA, software may qualify as a medical device when intended for diagnosing, treating, preventing, or mitigating disease.
  • The IMDRF defines Software as a Medical Device (SaMD) as software performing medical functions independently of a hardware medical device.
  • In Europe, EU MDR Annex VIII, Rule 11 governs software classification. Software supporting diagnostic or therapeutic decisions generally falls under Class IIa, IIb, or III, depending on the potential harm.
Medical Device vs. Non-Medical Software
Medical Device Not a Medical Device
App calculating personalized insulin doses Patient appointment scheduling app
AI detecting suspicious lesions in medical images General fitness and calorie tracker
Software analyzing ECG data for cardiac abnormalities Digital medical reference library without patient-specific recommendations

Medical Device Software Classification and Safety Classes

Medical device software is classified according to its potential risks, but different regulatory frameworks use different classification systems. IEC 62304 focuses on software safety, while the FDA and EU MDR classify medical devices based on regulatory risk. These classifications are related, but they are not interchangeable.

Understanding the differences helps determine which development processes, documentation, testing, and regulatory approvals your software will require.

Medical Device Software Classification: IEC 62304, FDA, and EU MDR
Framework Class Risk Level and Requirements
IEC 62304 A No injury or damage to health is possible from software failure. The standard’s Class A lifecycle requirements apply.
B Non-serious injury is possible as a result of software failure. Additional architecture documentation, integration testing, and risk controls.
C Death or serious injury is possible as a result of software failure. Most extensive design documentation, detailed verification, and traceability.
FDA I Low risk. General controls; most devices are 510(k)-exempt.
II Moderate risk. General and special controls apply; many devices require 510(k) clearance, although some are exempt.
III High risk. Generally requires Premarket Approval (PMA) and extensive safety and effectiveness evidence.
EU MDR I Lowest risk. Manufacturer self-declaration is generally possible.
IIa Moderate risk. Notified Body assessment generally required.
IIb Higher risk. More demanding conformity assessment and clinical evidence requirements.
III Highest risk. Strictest conformity assessment and clinical evidence requirements.

A device’s classification directly affects the development scope. Higher-risk software generally requires more extensive risk controls and verification evidence, while regulatory requirements may also increase the need for clinical evaluation and documentation.

These requirements influence project timelines and budgets. Identifying the appropriate classification early helps teams plan development activities and avoid costly changes later.

Why does the risk class affect the project? Risk classification directly affects development scope. Higher-risk software requires more detailed requirements, architecture records, risk management documentation, traceability, and verification activities. Testing must provide stronger evidence that the software performs safely under expected operating conditions and foreseeable failures.

These requirements also influence project timelines and budgets. Additional validation, regulatory reviews, clinical evidence, and documentation create more work before release. Identifying the appropriate classification early helps teams estimate development costs realistically, allocate specialist resources, and avoid expensive redesigns or submission delays.

Medical Device Software Development Standards and Regulations

Developing medical device software requires more than writing reliable code. Manufacturers must demonstrate that their products are safe, effective, traceable, and compliant throughout their lifecycle.

Different standards address different responsibilities: software development, quality management, risk control, usability, cybersecurity, and patient data protection. Requirements depend on the product’s intended use, target markets, and regulatory classification.

Key Medical Device Software Standards and Regulations
Standard or Regulation What It Covers When It Matters
IEC 62304 Software lifecycle, development, maintenance, and verification Throughout software development
ISO 13485 / FDA QMSR Quality management systems, design controls, and documented processes When establishing regulatory-compliant development and manufacturing
ISO 14971 Hazard identification, risk evaluation, and risk controls Throughout the device lifecycle
IEC 62366-1 Usability engineering and prevention of use-related errors When user interactions can affect safety
FDA Cybersecurity Guidance Secure design, vulnerability management, and Software Bill of Materials (SBOM) For cybersecurity risk and applicable FDA premarket submissions
EU MDR Device classification, clinical evaluation, conformity assessment, and post-market surveillance When placing medical devices on the EU market
HIPAA / GDPR Protected health information, privacy, and personal data processing When handling regulated health data
HL7 FHIR / DICOM / IEEE 11073 Healthcare data exchange, medical imaging, and device communication When integrating with EHRs, imaging systems, or connected devices

These frameworks serve different purposes. Regulatory obligations must be distinguished from technical standards and interoperability specifications. For example, HIPAA applies to covered entities and business associates, not automatically to every health app. Similarly, FHIR and DICOM support interoperability but do not independently establish regulatory compliance.

Recent Regulatory Updates for Medical Device Software

On February 2, 2026, the FDA’s Quality Management System Regulation (QMSR) took effect, incorporating ISO 13485:2016 by reference into 21 CFR Part 820. The FDA also updated its cybersecurity guidance in February 2026, clarifying recommendations for premarket documentation and software bills of materials (SBOMs).

For AI-enabled medical devices, the FDA’s final guidance on Predetermined Change Control Plans (PCCPs), published in August 2025, explains how manufacturers can document planned software modifications. Broader FDA guidance on AI lifecycle management remains in draft form.

In the EU, amendments to the AI Act moved the application date for high-risk AI systems covered by Article 6(1) and Annex I to August 2, 2028. This includes qualifying AI-enabled medical devices.

Medical Device Software Development Process

Medical device software development follows a structured lifecycle that addresses safety and regulatory requirements from the outset. The following seven stages outline a practical development process incorporating IEC 62304 lifecycle requirements alongside broader regulatory and validation activities. Risk management, documentation, and traceability continue throughout the entire lifecycle.

Medical device software development and integration

1. Intended Use and Risk Classification

Development begins by defining the software’s medical purpose, intended users, and clinical environment. These factors determine whether the product qualifies as a medical device and which regulations apply.

The team establishes the FDA or EU MDR risk classification, IEC 62304 software safety class, and potential regulatory pathway, such as 510(k), De Novo, or PMA. Early classification helps prevent costly changes later.

Output: Intended use statement, risk assessment, classification rationale, and regulatory strategy.

Common mistake: Starting with features before defining the software’s intended medical purpose.

2. Requirements and Architecture

Clinical needs are translated into measurable functional, safety, performance, and security requirements. Developers then design the software architecture, defining components, data flows, dependencies, and external interfaces.

Integration planning begins immediately, especially for connections with EHR systems, medical devices, or imaging platforms using standards such as HL7 FHIR, DICOM, or IEEE 11073. Requirements must remain traceable to design decisions, risk controls, and tests.

Output: Software Requirements Specification (SRS), architecture diagrams, interface specifications, and traceability matrix.

Common mistake: Leaving integration planning until the end of development.

3. Design and Usability

Medical software must be intuitive and safe in real clinical environments. Following IEC 62366-1, designers evaluate how users interact with the system and identify potentially harmful use errors. Wireframes and prototypes are tested with clinicians and other intended users. Feedback helps improve workflows, alerts, navigation, and critical information displays before coding begins.

Output: UX specifications, interactive prototypes, usability evaluation records, and use-related risk analysis.

Common mistake: Testing interfaces only with developers instead of actual healthcare professionals or other intended users.

4. Development

Engineers implement the approved architecture using secure coding practices, version control, automated checks, and peer code reviews.

Third-party components, including Software of Unknown Provenance (SOUP) and off-the-shelf (OTS) software, require documented evaluation. Teams must track dependency versions, known vulnerabilities, and changes affecting safety or security. Code changes remain linked to requirements and risk controls.

Output: Source code, code review records, dependency inventory, Software Bill of Materials (SBOM), and updated traceability records.

Common mistake: Failing to document third-party library versions and dependencies.

5. Verification and Validation

Verification confirms that the software meets specified requirements. Validation demonstrates that it fulfills its intended medical purpose and user needs. Testing includes unit, integration, system, regression, performance, and security testing.

Connected medical software should be evaluated using both simulators and representative physical devices to identify real-world communication and hardware issues. Test results, failures, and corrective actions must be documented.

Output: Verification and validation plans, test reports, defect records, and risk-control verification evidence.

Common mistake: Testing only with simulators rather than representative medical hardware.

6. Release and Regulatory Support

Before release, the team reviews completed testing, unresolved issues, risk controls, and software configuration. Regulatory documentation is prepared for applicable FDA submissions or EU MDR conformity assessments.

Medical software validation and regulatory compliance

Depending on the product, this may include software descriptions, architecture records, risk management files, and verification evidence. At this stage, medical device software development services may also involve preparing technical documentation to support regulatory submissions. Final release approval follows established quality management procedures.

Output: Release records, technical documentation, regulatory submission package, and approved software baseline.

Common mistake: Collecting regulatory evidence only after development is finished.

7. Post-Market Monitoring and Updates

The software lifecycle continues after deployment. Manufacturers monitor performance, investigate incidents, address cybersecurity vulnerabilities, and implement corrective actions. Maintenance also includes secure software and firmware updates, potentially delivered over the air (OTA).

Changes require impact assessments, appropriate regression testing, and updated documentation. A defined maintenance strategy helps ensure devices remain safe, reliable, and supportable throughout their operational life.

Output: Post-market surveillance records, maintenance plans, change assessments, and update validation reports.

Common mistake: Failing to plan secure software and firmware updates before product release.

If you plan a new product, our medical device software development services cover this process from discovery to support.

How Medical Devices Connect to EHR, Cloud, and Mobile Apps

Medical device integration connects healthcare equipment with mobile applications, cloud platforms, and electronic health records (EHRs). The goal is interoperability: enabling different systems to exchange, interpret, and use clinical data reliably.

Successful integration requires more than establishing a connection. Data must reach the correct patient record, retain its clinical meaning, and remain available despite network interruptions or device failures.

Medical device integration with cloud and EHR systems

The Connection Path

A typical medical device connectivity architecture consists of five layers:

Device → Gateway or Phone → App → Cloud → Hospital Systems (EHR)

  1. Device: Collects measurements such as blood pressure, glucose levels, ECG signals, or oxygen saturation.
  2. Gateway or Phone: Receives device data through BLE, USB, or other supported connections.
  3. App: Displays readings, validates incoming information, and manages user interactions.
  4. Cloud: Securely stores, processes, and synchronizes data while supporting remote access.
  5. Hospital Systems: Integrate validated data into EHRs, clinical dashboards, or monitoring platforms.

The architecture varies by product. Some devices communicate directly with cloud services or hospital systems without requiring every intermediate layer.

Connection Methods and Gateways

The connection method depends on device capabilities, power consumption, mobility, and clinical requirements.

Bluetooth Low Energy (BLE) is commonly used for wearable sensors and portable monitoring devices. It supports low-power communication with smartphones and dedicated receivers. Reliable BLE app development must account for pairing, reconnection, and data synchronization.

USB and serial interfaces remain relevant for bedside equipment, laboratory instruments, and legacy medical devices requiring wired connections.

Wi-Fi and cellular networks support remote monitoring and direct communication with hospital or cloud infrastructure, although coverage, security, and power consumption require consideration.

A medical device gateway may be needed when devices use incompatible protocols or cannot communicate directly with an EHR. In such cases, IoT integration services can help establish reliable data exchange between devices and healthcare systems. Gateways can also buffer measurements when connectivity is interrupted.

Data Standards

Choosing appropriate standards reduces custom integration work and helps systems interpret exchanged information consistently.

Healthcare Data Exchange and Interoperability Standards
Standard What It Does Where It Is Used Limitations
HL7 v2 / FHIR Exchange structured clinical information EHRs, hospital interfaces, healthcare APIs Version differences and local implementation requirements
DICOM Manages medical images and related metadata PACS, radiology, imaging devices Primarily focused on imaging workflows
IEEE 11073 Standardizes communication and data models for medical devices Patient monitors, connected health devices Device-specific profiles and implementation support vary
IHE Defines integration profiles using existing standards Cross-system clinical workflows Requires compatible profiles and coordinated implementation

No single standard guarantees interoperability. Systems must also agree on terminology, supported versions, security, and workflow behavior.

Typical EHR Integration Flow

Consider a connected blood pressure monitor. The device sends a reading to a mobile app or gateway, which checks the measurement, timestamp, unit, and device identifier.

An integration service then maps the reading to the EHR’s expected format, such as a FHIR Observation, and associates it with the correct patient using verified identifiers.

Before storing the measurement, the system checks for duplicates, missing information, and transmission errors. Failed deliveries enter a controlled retry queue, while acknowledgments confirm whether the receiving system accepted the message.

Audit logs record each transaction, making it possible to investigate missing readings or incorrect patient associations. Clinical availability should be confirmed at the receiving system, not assumed from successful transmission alone.

Common Integration Mistakes

Even well-designed integrations can fail because of overlooked operational details:

  1. Inconsistent measurement units: Missing conversions can produce incorrect clinical values.
  2. Connection loss: Network interruptions may cause missing or incomplete measurements.
  3. Firmware incompatibility: Device updates can change data formats or communication behavior.
  4. Missing retry mechanisms: Unacknowledged messages may be lost permanently.
  5. Unrealistic testing: Synthetic datasets may overlook real device errors and inconsistent patient records.
Vitaly TormanovVitaly TormanovHead of Mobile Solutions, SCAND

In medical software projects, the biggest challenges often emerge where different parts of the system interact. An application may work perfectly on its own, but connecting it to medical equipment or an existing healthcare platform introduces new risks. That’s why we pay particular attention to how data moves through the system and what happens when something doesn’t work as expected.

Cybersecurity and Patient Data

Medical device software often handles sensitive patient information and connects to hospital networks, mobile applications, and cloud platforms. Cybersecurity must therefore be addressed throughout the software lifecycle, not added shortly before release.

Security planning begins with threat modeling: identifying potential attack paths, vulnerable components, unauthorized access risks, and the consequences of compromised data or device functionality. These findings help teams prioritize appropriate security controls.

A Software Bill of Materials (SBOM) provides an inventory of software components and dependencies, making it easier to identify known vulnerabilities. Regular security assessments, vulnerability monitoring, and controlled software updates help address emerging threats after deployment.

Common security measures include encryption of data in transit and at rest, secure authentication, role-based access control (RBAC), and audit logging. For connected devices, update mechanisms should also verify software authenticity and support safe recovery from failed installations.

Patient data protection introduces additional legal obligations. In the United States, HIPAA applies to covered entities and business associates handling protected health information. In the European Union, GDPR governs the processing of personal data, including health information, subject to its scope and applicable legal requirements.

Implementing security controls does not automatically establish regulatory compliance. Requirements depend on the product, data flows, organizational responsibilities, and jurisdictions involved. Security, privacy, and regulatory assessments should be documented and reviewed throughout development and maintenance.

AI in Medical Device Software

Artificial intelligence is increasingly used in medical devices for image analysis, patient monitoring, anomaly detection, and clinical decision support. AI medical device software development requires careful consideration of where models operate, how their performance is validated, and how they are updated after release.

AI integration into medical device software

AI models can run in three environments:

  • On-device (edge AI): Processes data locally, reducing latency and dependence on internet connectivity.
  • Gateway or smartphone: Analyzes device data before forwarding results to other systems.
  • Cloud: Supports computationally intensive models, centralized processing, and integration with healthcare platforms.

The right architecture depends on processing requirements, connectivity, privacy, and clinical risks. AIoT development can combine connected medical devices with AI-powered data processing.

Model validation is essential. Developers must evaluate accuracy, sensitivity, specificity, and performance across representative patient populations and clinical conditions. Testing should also address biased datasets, unexpected inputs, and performance degradation over time.

AI models may need updates as new data becomes available. The FDA’s Predetermined Change Control Plan (PCCP) guidance explains how manufacturers can describe planned modifications, validation methods, and impact assessments in a marketing submission. An FDA-authorized PCCP can allow specified changes without a new submission for each modification.

For AI systems intended to support clinical decisions, developers should clearly define the role of healthcare professionals and communicate the model’s limitations. The required level of human oversight depends on the software’s intended use and applicable regulations.

What Affects the Cost of Medical Device Software Development

The cost of medical device software development depends on more than the number of features. Regulatory classification, integration complexity, testing requirements, and long-term maintenance all influence the overall budget.

Unlike conventional applications, medical device software requires documented evidence of safety and performance. The higher the risk and technical complexity, the more resources may be needed for development, verification, and regulatory preparation.

Key Factors Affecting Development Costs
Factor How It Affects Cost
Risk classification Higher-risk devices generally require more extensive safety controls, testing, and regulatory documentation.
Feature complexity Real-time monitoring, clinical algorithms, AI, and advanced analytics increase engineering and validation work.
Number of integrations Connecting multiple devices, EHRs, and cloud platforms adds interoperability and testing requirements.
Supported platforms Developing for mobile, web, desktop, or embedded systems increases implementation and compatibility testing.
Documentation and V&V Requirements traceability, risk management, verification, and validation require dedicated engineering and quality resources.
Maintenance and support Cybersecurity updates, vulnerability monitoring, firmware compatibility, and post-market activities create ongoing costs.

How Development Costs Are Estimated

A reliable estimate is prepared after the discovery phase, when intended use, regulatory classification, technical requirements, integrations, and verification scope have been assessed.

The engagement model also influences how resources and responsibilities are organized.

Medical Device Software Development Engagement Models
Engagement Model Best Suited For
Dedicated team Long-term development, product evolution, and ongoing maintenance
Project-based development Projects with clearly defined scope, deliverables, and responsibilities
Staff augmentation Extending an existing team with specialized engineering or QA expertise

The right approach depends on project maturity, internal capabilities, and the level of control required. A discovery-based estimate helps establish a realistic budget without relying on generic pricing ranges that may not reflect the product’s regulatory or technical complexity.

How to Choose a Medical Device Software Development Partner

Experience in healthcare software doesn’t necessarily mean a company has worked on regulated medical devices. Developing a patient portal involves different requirements than building software that controls medical equipment or supports diagnosis. When evaluating a partner, focus on experience relevant to your product’s intended use and regulatory classification.

Secure medical software development and collaboration

Check Relevant Experience

Ask potential partners to describe a comparable project and explain their responsibilities. Did they work within the manufacturer’s quality management system (QMS)? Did they prepare verification evidence or support regulatory submissions?

These details reveal more than a portfolio of healthcare applications. They also help establish which activities the provider can handle and which remain the manufacturer’s responsibility.

Evaluate Risk Management Practices

Medical device development requires documented evidence that risks have been addressed. A practical way to assess a team is to ask how it would handle a change to a safety-critical feature after testing has begun.

The answer should explain how the change affects existing risk controls and which tests or documentation need updating. This reveals more about the team’s experience than simply asking whether it follows IEC 62304 or ISO 14971.

Assess Integration Capabilities

For connected medical devices, knowledge of HL7 FHIR or DICOM is not enough. Ask how the team tests communication with physical equipment and handles unexpected device behavior.

Relevant experience should include testing with representative hardware and verifying that clinical data reaches the correct patient record.

Questions to Ask Before Hiring

  • Have you worked on devices with similar intended uses or risk classifications?
  • How do you manage changes affecting software safety?
  • What regulatory documentation can your team prepare?
  • How do you test integrations with physical medical devices?

A suitable partner should support its answers with project examples and clearly define its responsibilities before development begins.

SCAND Experience in Healthcare and Device Software Development

SCAND brings experience from both healthcare applications and hardware-related software projects. This combination is particularly relevant when developing connected solutions that require reliable communication between devices and applications. Selected projects from our portfolio include:

  • React Native Healthcare App for a Private Medical Practice: A private clinic needed a way to manage appointments and offer remote consultations. Built with React Native and Node.js, the app uses Twilio for video calls and Google Calendar to avoid scheduling conflicts. Development took 10 weeks.
  • Data Acquisition Software for Scientific Sensors: SCAND developed a Flutter application that collects measurements from scientific instruments via Bluetooth or USB and displays real-time graphs. The solution runs on mobile and desktop platforms and can be used in medical research.
  • Custom AOSP App for a Biometric Scanner: The project involved processing synchronized RGB and infrared video streams on resource-constrained hardware. SCAND used Kotlin, C/C++, and FFmpeg to build the application, which also supports interrupted file transfers and remote updates.

Together, these projects demonstrate SCAND’s ability to work across application and device-level development. For clients building connected products, this experience is valuable when software must operate within hardware constraints rather than in a conventional application environment.

Custom healthcare software development services

Frequently Asked Questions (FAQs)

What is medical device software development?

Medical device software development involves creating software embedded in medical devices, standalone Software as a Medical Device (SaMD), or applications connecting medical equipment to other systems. It follows a risk-based lifecycle covering requirements, testing, regulatory documentation, and post-market maintenance, guided by standards such as IEC 62304.

What is the difference between SaMD and SiMD?

SaMD (Software as a Medical Device) performs a medical function independently of medical hardware, such as software analyzing medical images. SiMD (Software in a Medical Device) operates as part of a physical medical device, such as infusion pump firmware. Both may be subject to medical device regulations, but their development and verification requirements differ.

What are the stages of medical device software development?

The seven main stages are intended use and risk classification, requirements and architecture, design and usability, development, verification and validation, release and regulatory support, and post-market monitoring. Each stage produces documentation supporting safety, traceability, and regulatory assessments.

Does medical device software need FDA clearance?

Not always. Requirements depend on the software's intended use, device classification, and applicable exemptions. Some devices require 510(k) clearance, others follow the De Novo classification pathway or require Premarket Approval (PMA). Certain lower-risk software functions may be exempt from premarket submission requirements.

What is IEC 62304?

IEC 62304 is an international standard defining software lifecycle processes for medical devices, including development, maintenance, and risk-related activities. It establishes three software safety classes: A, B, and C, with increasing process and documentation requirements depending on potential harm.

How do medical devices integrate with EHR systems?

Medical devices typically transmit data through a gateway, mobile app, or integration platform. The information is then mapped to supported standards such as HL7 v2 or FHIR and transferred to the EHR. Reliable integration requires patient matching, data validation, duplicate detection, error handling, and testing with representative device data.

How much does medical device software development cost?

Costs depend on risk classification, functionality, supported platforms, integrations, regulatory documentation, and verification requirements. A discovery phase provides the foundation for a realistic estimate by clarifying technical scope, compliance needs, and development responsibilities. Generic pricing ranges rarely reflect the complexity of an individual medical device software project.

Author Bio
Head of Mobile Solutions
Vitaly Tormanov Head of Mobile Solutions
Vitaly Tormanov is a seasoned software architect and Head of Mobile Solutions Department at SCAND, with nearly 20 years of professional experience in software development.

Looking for a Custom Fix?

SCAND’s the company to call for smart solutions and easy-going consulting.

Shoot us a message
and let's get started!
Contact us
Need Mobile Developers?

At SCAND you can hire mobile app developers with exceptional experience in native, hybrid, and cross-platform app development.

Mobile Developers Mobile Developers
Looking for Java Developers?

SCAND has a team of 50+ Java software engineers to choose from.

Java Developers Java Developers
Looking for Skilled .NET Developers?

At SCAND, we have a pool of .NET software developers to choose from.

NET developers NET developers
Need to Hire Web Developers Faster?

Bring the right skills to your project from day one.

Web Developers Web Developers
Need to Staff Your Team With React Developers?

Our team of 25+ React engineers is here at your disposal.

React Developers React Developers
Searching for Remote Front-end Developers?

SCAND is here for you to offer a pool of 70+ front end engineers to choose from.

Front-end Developers Front-end Developers
Other Posts in This Category
View All Posts

This site uses technical cookies and allows the sending of 'third-party' cookies. By continuing to browse, you accept the use of cookies. For more information, see our Privacy Policy.